Security:
Anti-shock types: Facilities I, no internal power supply Anti-electric Shock Degree: B Working Voltage: AC 100 V~240 V Frequency: 50 Hz/60 Hz P<60 VA Fuse: T1.6AL250VDisplay
Dimensions: 8.0 “color LCD display, folding 60 degree
Display
Content: bed No. , pregnancy weeks, age, paper speed, date, time , volume, alarm status, transducer connection status, recorder status, FHR data and wave, Contraction data and wave, Fetal move times and mark etc.
Print:
Record Paper two-double type Z
Print Width: 112 mm
Valid Print Width: 104 mm
Paper output speed: 1 cm/min、2 cm/min、3 cm/min(optional)
Data Precision: ±5 %(X Roll),±1 %(Y Roll)
Record Content: hospital , bed No. ,name, pregnancy weeks, patient No., paper speed, date, time , FHR data and wave, Contraction data and wave, Fetal move times and mark etc.
Ultrasound probe:
Nominal Frequency: 1.0 MHz
Work Frequency: 1.0 MHz±10 %
Negative peak sound pressure : p_<1 MPa
Output beam intensity : Iob<20 mW/cm2
The peak time space peak intensity: Ispta<100 mW/cm2
The average time space peak intensity: Isata<10 mW/cm2
FHR Rang: 50 BPM~240 BPM
Resolution: 1 BPM
Accuracy : ±2 BPM
TOCO
TOCO range: 0~100 %
Resolution: 1 %
Nonlinear error: <±10 %
RZ way: Manually
Fetal Marking
For the manual button (the operation of pregnant women), there will be a mark display in the bottom area of FHR wave display section.
FHR Alarm:
Alarm for high and low FHR, which exceeds appointed limit.